Ceramic durability

By using oxide ceramics, KROHNE employs a technically superior material for electromagnetic flowmeters (EMF) and pressure-measuring devices. This material offers long-term resistance to corrosive and abrasive media and does not leach. Additionally, these ceramics are resistant to temperature shocks and completely safe against gas permeation and leakage – providing a unique combination of properties for high-performance applications across all industrial sectors.

Highlights

  • High-resistance to abrasion

  • Suited for pastes and sludges with very high solid content

  • The helium leakage rate of ceramic including electrodes is less than 10-7 (hPa * l)/s, this means „gas-tight“

  • Vacuum-resistant up to 0 hPa

  • High mechanical strength and form stability

  • 0 % porosity (important for aggressive outgassing or toxic media)

  • Excellent surface quality and very low roughness (Ra < 0.8 µm)

  • Over 30 years of experience with industrial ceramic

  • Successful adaptions for various applications

Overview of KROHNE devices with oxide ceramics:

Flowmeters

Installation

Size/DN

Ceramic

Electrode

OPTIFLUX 5000

Flange

15 ... 300

FZM

CERMET, sintered-in

Sandwich

2.5 ... 15

FZM

CERMET, sintered-in

Sandwich

25 ... 100

FZT

Platinum pin electrode with plated head, pin sintered-in

BATCHFLUX 3100

Sandwich

2.5 ... 40

FZM

Platinum pin electrode with plated head, pin sintered-in

BATCHFLUX 5500

Sandwich

15

FZM

CERMET, sintered-in

OPTIFLUX 7000

Flange

2.5 ... 40

FTM

Capacitive, on ceramic tube

Sandwich

25

FZM

Capacitive, on ceramic tube

40 ... 100

FZT

Capacitive, on ceramic tube

Ceramics used:

FZT: FRIALIT® FZT: ZrO2-stabilised Al2O3, Bending strength = 460 MPa*, E-module = 360 GPa

FZM: FRIALIT® FZM: MgO partially stabilised ZrO2, Bending strength = 500 MPa*, E-module = 200 GPa

*1 MPa = 1 N/mm2

Pressure instruments

OPTIBAR PC 5060 Pressure transmitter

OPTIBAR­LC 1010 Immersion probe

Measuring ranges

0.025…100 bar

0...1 to 0...100 mH2O

Reference accuracy

<0.05 to 0.2% of URV

<±0.35% FSO

Process temperature

–40...+150°C

-40...+80°C

Process connections / material

Thread of more than 1/4¨, ASME flange, EN, DIN, JIS from DN15 in 316L, C276, Duplex, PVDF

Thread R1¨,

Flange DN50,

cable: TPE

Approvals

Ex ia, Ex d, SIL2/3

ATEX/IECEx Ex ia 1G,

drinking water

approvals (KTW, ACS)

Ceramics used:

Sapphire: 99.96 % Al2O3 ceramic, Bending strength = 630 MPa, E-module¬=¬406 GPa

Special features of flowmeters with ceramic liner:

  • Oxide ceramic is very resistant to acids and lyes. (Nitric acid and nitrohydrochloric acid or even hot concentrated hydrochloric acid cannot damage the oxide ceramic)

  • Adherence to the pressure equipment directive (DGLR)

  • Resistant to abrasive media: long service life especially compared to plastic liners

  • Fully vacuum-resistant

  • Dimensionally stable: high resistance to temperature shocks of up to 120 K and pressure stresses

  • Very high burst pressure/overload resistanct

  • No deposits/gumming on surface, diffusion tight and immune to leakage

  • Sintered-in CERMET electrodes

  • Leading metrological institutes use KROHNE ceramic EMFs as their master/ reference meter

Special features of pressure measuring devices with ceramic diaphragm

  • Suitable for pressure applications of up to 100 bar and 150 °C

  • Overload-proof up to 200 times the nominal measurement range

  • 100 % hydrogen resistance

  • 100 % diaphragm breakage detection

  • In the event of diaphragm breakage this is detected by the sensor electronics

  • Ten times harder than stainless steel

  • Absolutely vacuum-tight

관련 제품

다양한 용기 크기의 까다로운 체적 충전 어플리케이션을 위한 전자 유량계

  • 우유, 고체 함량이 있는 과일 음료, 세제 등 전도성 액체의 냉온 충전은 물론 공정 배치에서 고정밀 측정이 가능합니다.
  • 최대 16 bar용 세라믹 측정 튜브 및 소결 전극 포함
  • 플로우 튜브는 EC1935/2004, FDA, GB 4806, 3A 인증 최대 DN25 / 1" 준수
  • 주파수 출력 통합

일반적인 용기 크기의 체적 충전을 위한 전자 유량계

  • 전도성 액체(예: 물, 탄산음료 등) 충전용 시장에서 가장 컴팩트한 충전 마그네틱 계량기.
  • 가볍고 경제적인 디자인, 최대 8 bar 용 세라믹 유량관 포함
  • DN15 / ½" 세라믹 튜브 는 EC1935/2004, FDA, GB 4806을 준수합니다
  • 주파수 출력 통합

IFC 100 및 IFC 300 신호 변환기와 결합하기 위한 전자 유량 센서

  • 전체 보어 설계: 제한 없는 파이프 단면적
  • 연결 크기: DN2.5…300 / ⅒…120"
  • 튜브 재질: 세라믹

기본 압력 및 레벨 어플리케이션을 위한 세라믹 다이어프램이 장착된 압력 트랜스미터

  • 뛰어난 과부하 저항, 장기적인 안정성내식성을 갖춘 초소형 디자인
  • 정전식 측정 셀 충전액 미포함: 매립형 또는 전면 플러시형 세라믹 다이어프램
  • 측정 범위: 0.1...40 bar,최대 +125°C
  • 나사산: G¼, G½, G¾(수형), 금속이 없는 PVDF 또는 PP-HT로도 제공

수질 및 폐수 어플리케이션을 위한 수중 레벨 프로브

  • 분지 및 우물의 연속적인 레벨 측정용
  • 부식에 강한 세라믹 측정 셀 사용
  • 수위(물): 1...100 m / 3.3...328 ft; 압력: 0.1...10 bar / 1.45...145 psi
  • 2-wire, 1 x 4...20 mA, HART®

레벨, 차압 및 밀도 측정을 위한 전자식 차압 트랜스미터

  • 통합 전자 차압 측정
  • 긴 모세관 라인이 있는 다이어프램 씰 설치의 대안
  • 추가 공정 정보를 위한 통합 정압 측정
  • 2-wire, 4...20mA, HART®, FF, Profibus-PA, Bluetooth®